4.4 Setup
INFORMATION
The distances between camera stands and the force measurement plate given in this document provide an
approximate indication of space requirements and assist with setup. Subsequently the camera stands and force
measurement plate are aligned with each other using crosshairs in the app (section "Alignment of the stands and
force measurement plate" – see Page 83). Deviations from the distance and angle values specified for the setup
are possible.
~90°
~800 mm
~90°
743L500 3D L.A.S.A.R. Posture
~700 mm
~90°
► Note the following points to select a suitable setup
location:
→ Find a firm, level surface (such as laminate,
concrete, PVC flooring, short pile carpet).
→ Position the stands as close to the walls as pos
sible so the cable of the sagittal camera stand
can be laid close to the wall, avoiding the risk of
tripping.
→ Windows or light sources that are captured by
the cameras must have a means of shading
them (for instance roller shutters, curtains,
switching off) if the shutter priority of the camer
as is negatively influenced by backlighting.
► CAUTION! Proceed carefully during setup to
avoid injury and damage to the product com
ponents (such as cables).
INFORMATION: The illustration shows the
basic setup with the sagittal cameras aligned in
the upper case section on the patient's right
side. There are two possibilities to capture the
patient's left sagittal side: 1. The patient turns
by 180° to stand on the force measurement
plate. 2. The force measurement plate is
rotated 90° in place and then calibrated again.
Alternatively the upper section of the case can
be aligned with the patient's right side for the
basic setup.
Set up the lower and upper sections of the case,
the force measurement plate and the camera
stands, and connect the cables.
Preparation for use
75