5.4.7 Camera connection failed
Camera connection failed
Please check the cables between the PC and the
cameras
0.0 Nm
0.0 %
Camera connection failed
Please check the cables between the PC and the
cameras
5.4.8 Use outside of the permissible temperature range
Use outside the permissible
temperature range
The operating temperature is outside
the permissible temperature range.
Measurement errors may occur as a result.
743L500 3D L.A.S.A.R. Posture
Live view
OK
0.0 Nm
0.0 %
1
2
OK
Live view
Use outside the permissible
temperature range
The operating temperature is outside
the permissible temperature range.
Measurement errors may occur as a result.
OK
0.0 Nm
0.0 Nm
0.0 %
0.0 %
1
2
OK
Error message text:
: "Camera connection failed"
: "Please check the cables between the PC and the
cameras"
: "OK"
Possible causes:
•
USB plug not connected to mini-computer
•
Mini-computer software has crashed
•
Defective cable
•
Defective camera
Measures:
► Carry out the appropriate measures depending on
the cause (for example connect USB plug; restart
the mini-computer; have defective components
repaired by Ottobock Service).
► 1x tap on "OK".
Error message text:
: "Use outside of the permissible temperature range"
: "The operating temperature is outside the permiss
ible temperature range."
: "Measurement errors may occur."
: "OK"
Possible causes:
•
The ambient temperature is < +15 °C.
The ambient temperature is > +30 °C.
•
Measures:
► Use the product exclusively within the permissible
temperature range (+15 °C – +30 °C).
► Tap on "OK" once.
Use