Error while charging the product
LED on
LED on bat­
power
tery charger
supply
Error after disconnecting the battery charger (error on self-test)
Beep
Vibration
signal
signal
3x short
68 | Ottobock
Error
Country-specific plug adapter not
fully engaged on power supply
Non-functional socket
Defective power supply
No connection between battery
charger and power supply
Defective battery charger
Battery is fully charged (or con­
nection with product is interrup­
ted).
Error
Minor error:
e.g. service interval has been
exceeded, disruption of a sensor
signal
Resolution
Check whether the country-spe­
cific plug adapter is fully
engaged on the power supply.
Check socket with another elec­
tric appliance.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Check whether the charging
cable plug is fully engaged on the
battery charger.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Take note of the beep signals to
differentiate between these
states.
When the battery charger is con­
nected or disconnected, a self-
test is conducted and confirmed
by a single beep and vibration
signal.
The battery is fully charged if this
beep signal is heard.
If no beep signal is heard, the
connection to the product is
interrupted.
If the connection to the product is
interrupted, an authorised
Ottobock Service Centre must
inspect the product, battery char­
ger and power supply.
Resolution
Contact Ottobock