13.2.3 Error messages while establishing a connection with the cockpit app
Error message
Device not found
Component was con
nected to another
device. Establish con
nection?
13.2.4 Status signals
Battery charger is connected
LED on
LED on bat
power
tery charger
supply
Battery charger disconnected
Beep
Vibration
signal
signal
1x short
1x short
Battery charge level
Battery char
ger
Battery is being charged, battery charge level is less than 50%
Battery is being charged, battery charge level is over 50%
Battery is fully charged (or connection with product is interrupted).
Take note of the beep signals to differentiate between these states.
When the battery charger is connected or disconnected, a self-test is conduc
ted and confirmed by a single beep and vibration signal.
The battery is fully charged if this beep signal is heard.
If no beep signal is heard, the connection to the product is interrupted.
Cause
The connection could
not be established as no
component with the
entered serial number
could be found.
The component was
connected to another
device/remote control
A current connection to
the prosthesis was inter
rupted
Event
Power supply and battery charger operational
Event
Self-test completed successfully. Product is operational.
Correction
Compare the serial number entered with
that of the component, then try to establish
the connection again.
To disconnect the original connection, tap
the "OK" button.
If the original connection is not to be dis
connected, tap the "Cancel" button.
Check the following points:
•
Distance from the prosthesis to the
device
•
Battery charge level of the prosthesis
•
Bluetooth on the prosthesis switched
on? (see Page 60)
•
If multiple prostheses were stored in
the remote control, was the correct
prosthesis selected?
Ottobock | 69