Masimo RD SET Serie Manual Del Usario página 5

Ocultar thumbs Ver también para RD SET Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 20
High oxygen concentrations may predispose a premature infant to retinopathy. Therefore, the upper alarm limit for the oxygen saturation
must be carefully selected in accordance with accepted clinical standards.
• Caution: Replace the sensor when a replace sensor message is displayed, or when a low SIQ message is consistently displayed after
completing the low SIQ troubleshooting steps identified in the monitoring device operator's manual.
• Note: The sensor is provided with X-Cal™ technology to minimize the risk of inaccurate readings and unanticipated loss of patient
monitoring. The sensor will provide up to 168 hours of patient monitoring time or up to 336 hours for sensors with a replaceable tape. After
single-patient use, discard sensor.
INSTRUCTIONS
A) Site Selection
• Always choose a site that is well perfused and will completely cover the sensor's detector window.
• Site should be cleaned of debris and dry prior to sensor placement.
RD SET Adt: Adult Sensor
> 30 kg The preferred site is the middle or ring finger of non-dominant hand.
RD SET Pdt: Pediactric Sensor
10–50 kg The preferred site is middle or ring finger of non-dominant hand.
RD SET Inf: Infant Sensor
3–10 kg The preferred site is the great toe. Alternatively, the toe next to the great toe, or the thumb can be used.
10–20 kg The preferred site is the middle or ring finger of the non-dominant hand.
RD SET Neo: Neonatal/Adult Sensor
< 3 kg The preferred site is the foot. Alternatively, across the palm and back of the hand can be used.
> 40 kg The preferred site is the middle or ring finger of non-dominant hand.
RD SET NeoPt/NeoPt-500: Preterm Sensors
< 1 kg The preferred site is the foot. Alternatively, across the palm and back of the hand can be used.
B) Attaching the sensor to the patient
1. Open the pouch and remove the sensor. Remove the backing from the sensor, if present.
Atd sensor for ADULTS (> 30 kg) and Pdt sensor for PEDIATRICS (10–50 kg)
2. Refer to Fig. 1a. Orient the sensor so that the detector can be placed first. Place the tip of the finger on the dashed line with the fleshy part of
the finger covering the finger outline and detector window.
3. Refer to Fig. 1b. Press the adhesive wings, one at a time, onto the finger. Complete coverage of the detector window is needed to ensure
accurate data.
4. Refer to Fig. 1c. Fold the sensor over the finger with the emitter window (
time, around the finger.
5. Refer to Fig. 1d. When properly applied, the emitter and detector should be vertically aligned (the black lines should align). Reposition if
necessary.
Inf sensor for INFANTS (3–10 kg)
2. Refer to Fig. 2a. Direct the sensor cable so that it runs along the top of the foot. Position the detector on the fleshy pad of the great toe.
Alternatively, the toe next to the great toe, or the thumb can be used (not shown).
3. Refer to Fig. 2b. Wrap the adhesive wrap around the toe so the emiter is positioned on the nailbed of the great toe. Complete coverage of the
detector window is needed to ensure accurate data.
4. Refer to Fig. 2c. Ensure that the emitter window (
reposition if necessary.
Neo sensor for NEONATES (< 3 kg) and NeoPt/NeoPt-500 sensor for PRETERMS (< 1 kg)
2. Refer to Fig. 3a. For fragile skin, the stickiness of the medical grade adhesive can be diminished or eliminated by daubing the adhesive areas
with a cotton ball or gauze.
3. Refer to Fig. 3b. Direct the sensor cable toward the ankle (or wrist). Apply the sensor around the lateral aspect of the foot (or hand), aligned
with the fourth toe (or finger). Complete coverage of the detector window is needed to ensure accurate data.
4. Refer to Fig. 3c. Wrap the adhesive/foam wrap around the lateral aspect of the foot (or hand) and ensure that the emitter window (
directly opposite of the detector. Be careful to maintain proper alignment of the detector and emitter windows while attaching adhesive/foam
wrap to secure the sensor.
5. Refer to Fig. 3d. Verify correct positioning and reposition if necessary.
Neo sensor for ADULTS (> 40 kg) Inf Sensor for INFANTS (10–20 kg)
2. Refer to Fig. 4a. Direct the sensor cable so that it runs along the top of the hand. Position the detector on the fleshy part of the finger.
Alternatively, the sensor may also be applied to the toe (not shown).
3. Refer to Fig. 4b. Wrap the adhesive wrap around the finger so the emitter window (
detector. Complete coverage of the detector window is needed to ensure accurate data.
4. Refer to Fig. 4c. Check the sensor to verify correct positioning and reposition if necessary.
C) Attaching the Sensor to the Patient Cable
1. Refer to Fig. 5a. Orient the sensor's connector tab so that the side with the "shiny" contacts is facing up. Orient the patient cable with the color
bar and finger grips facing up.
2. Refer to Fig. 5b. Insert the sensor tab into the patient cable until there is a tactile or audible click of connection. Gently tug on the connectors
to ensure a positive contact. Tape may be used to secure the cable to the patient for ease of movement.
D) Reattachment
• The sensor may be reapplied to the same patient if the emitter and detector windows are clear and the adhesive still adheres to the skin.
• If the adhesive no longer adheres to the skin, use a new sensor.
NOTE: When changing application sites, or reattaching sensor, first disconnect the sensor from the patient cable.
E) Disconnecting the Sensor from the Patient Cable
1. Refer to Fig. 6. Pull firmly on the sensor connector to remove it from the patient cable.
NOTE: To avoid damage, pull on the sensor connector, not the cable.
) positioned over the fingernail. Secure the wings down, one at a
) aligns on the top of the toe directly opposite the detector. Verify correct positioning and
5
) aligns on the top of the finger directly opposite the
) aligns
9028F-eIFU-1117

Publicidad

Tabla de contenido
loading

Tabla de contenido