English
2. Clean foot pedal by wiping with non-abrasive cloth and neutral pH
detergent and deionized, distilled, or purified water after each case.
3. Dry thoroughly with non-abrasive cloth.
Dissection Tools
Warning: Do not reprocess dissection tools. Dissection tools are
disposable and SINGLE USE ONLY.
Dispose of properly in sharps container.
Note: The clinical processing instructions provided have been validated
by DePuy Synthes Power Tools for preparing a non-sterile medical device;
this instruction is provided in accordance with ISO 17664 and ANSI/AAMI
ST81.
Mechanical/Automated Cleaning
Warnings:
• Manual pre-cleaning must be completed prior to mechanical
cleaning.
• Cannulations must be pre-cleaned with appropriate brushes as
described in pre-cleaning instructions.
• At the point of use the device must be cleaned as soon as possible
after use to prevent drying of blood, tissue, other biological debris
and contaminants on the device.
• National regulations must be observed.
• You must follow standard hospital procedures.
• You must follow manufacturer's recommendations for any
detergents, disinfectants and washing equipment used.
• Do not reprocess dissection tools. Dissection tools are disposable
and SINGLE USE ONLY.
Cautions:
• Rx Only: United States Federal law restricts this device to sale by or
on order of a physician or other licensed healthcare provider.
• Only the devices listed in Table 1 may be reprocessed using the
mechanical/automated cleaning method.
• Do not use any alkaline detergent with a pH greater than 11. Alkaline
detergents with higher pH attack grease and seals which can
increase wear and cause device to malfunction.
• Do not use ultrasonic equipment.
• Do not use hypochlorite solution (i.e. bleach).
6