Note: Brushes may be inserted from distal or proximal end of
attachments and bearing sleeves.
Warning: ACB and SDCB are single use and should only be used to clean
one set of attachments.
Caution: For MA-D20-G1 and Minimal Access Attachment Bearing
Sleeves use SDCB only. Do not force ACB or SDCB into or through
attachments or bearing sleeves. Do not use pipe cleaners.
Figure 1
4. Rinse Thoroughly
• Rinse device thoroughly using running hot deionized, distilled, or
purified water for a minimum of 2 minutes.
• Use a syringe or pipette filled with hot deionized, distilled or purified
water to flush lumens or channels.
• Manipulate any moving parts such as release sleeves (Figure 1).
Caution: Do not use high pressure water or air. Do not rinse attachment
with saline solution.
5. Inspect
• Visually inspect attachment for visible soil. Repeat Steps 2 thru 5 if
visible soil remains.
6. Lubricate
Caution: This step is for attachments only. Do not lubricate handpieces.
Option 1: 05.001.078, Lubricant for Anspach Systems, 110 ml
Warning: Do not use Gravity Air Displacement Steam Autoclave
Sterilization Process.
• For the following attachments apply one (1) spray pump spray to the
proximal end of the attachment.
SHORT-G1
SHORT-HD-G1
MEDIUM-G1
MEDIUM-HD-G1
4
Release Sleeve
LONG-G1
LONG-HD-G1
MIA16-G1
CRANI-A-G1
CRANI-L-G1
CRANI-P-G1
MA-10S
MA-10C
MA-15S
MA-15C
MA-15ST
MA-19ST