• Handpieces must not be immersed in cleaning solution but
may be sprayed with it.
• Fully demineralized (deionized, distilled, or purified) water
must be used during final rinse to prevent corrosion and
spots.
Note: Use of alkaline detergents can cause color to fade but this
does not impair function of device.
Table 1: List of Devices
Device Catalog Number
EG1A
SHORT-G1
SHORT-HD-G1
MEDIUM-G1
MEDIUM-HD-G1
LONG-G1
LONG-HD-G1
MIA16-G1
QD8-G1
QD8-S-G1
QD11-G1
*MA-10S, MA-10C, MA-15S, MA-15C, MA-15ST, MA-19ST.
Warning: Consult G1 Cleaning and Sterilization Basket Loading
Diagram for validated loading configurations.
Note: MA Bearing Sleeves are compatible with MA-D20-G1,
MA-D20, MA-DRIVER and MA-D20BM.
Operating Room (OR) Personnel Instructions
1. After completion of procedure, remove external debris by wiping
attachments and handpiece.
2. Properly dispose of open or used dissection tools.
Reprocessing Personnel Instructions
Manual Pre-Cleaning of Handpiece
Note: The instructions in this section are specific to the eG1
Electric Handpiece.
1. Preparation
• Assemble all necessary supplies including Attachment Cleaning
Brush (ACB), sponge, lint-free cloth and soft bristle brush
Device Catalog Number
QD11-S-G1
QD14-G1
QD14-S-G1
OCM-G1
MA-D20-G1
MA Bearing Sleeves*
CRANI-P-G1
CRANI-A-G1
CRANI-L-G1
CSR60-G1
7