Sterilization
Before sterilizing:
• Ensure sterilization equipment is in proper working order as
specified by manufacturer.
• Assure equipment manufacturer's instructions are properly
employed by trained and qualified personnel.
• Assure actual cycle employed has been properly validated
for the device(s)/load configuration being processed and
appropriate sterilization indicator devices are included for
each process and cycle.
Note: Prior to using G1 Cleaning and Sterilization Basket for first
time, attach connector sleeve to basket by clipping onto internal
support structure (Figure 2). The connector sleeve should remain
with basket for future use.
Warning: Devices must be cleaned prior to sterilization using the
validated cleaning method outlined in these instructions.
1. Loading the Basket
• Load devices in G1 Cleaning and Sterilization Basket by
securing them in appropriate brackets, as shown in G1
Cleaning and Sterilization Basket Loading Diagram, with
distal end angled down where applicable.
• Remove connector sleeve from handpiece connector (Figure 1).
Replace connector sleeve in its bracket and lay handpiece
connector in bottom of basket.
Warning: Handpiece connector sleeve must be removed from the
handpiece connector prior to sterilization (Figure 1). Consult G1
Cleaning and Sterilization Basket Loading Diagram for validated loading
configurations. Angled attachments must be loaded in the correct
orientation as shown in Figure 2.
Figure 1
2. Packaging
• When wrapping the basket the lid must be used.
Figure 2
13