Caution: Connector end of electric handpiece must be sealed with
connector sleeve provided with basket (Figure 3).
Warning: Consult G1 Cleaning and Sterilization Basket Loading
Diagram for validated loading configurations. Angled attachments
must be loaded in the correct orientation as shown in Figure 4.
Figure 2
Figure 4
2. Mechanical/Automated Cleaning Cycle Process Parameters
Cycle
Rinse
Pre-wash
Cleaning
Rinse
Thermal
Processing
Dry
Caution: Do not use neutralizing agent.
3. Inspect
• Visually inspect devices for soil. Repeat manual pre-clean
and mechanical/automated clean if visible soil remains.
• Remove connector sleeve from connector end of handpiece
and drain until no visible droplets are coming from the
handpiece and handpiece connector.
Figure 3
Duration
(minimum)
2 minutes
1 minute
2 minutes
5 minutes
5 minutes
40 minutes
Cleaning Instructions
Cold deionized (DI) or
purified water (PURW)
Warm DI or PURW water
(≥ 40°C); use detergent
Warm DI or PURW water
(≥ 45°C); use detergent
Rinse with DI or PURW
water
Hot DI water, ≥ 93°C
≥ 90°C
11