English
• Remove attachment and dissection tool from handpiece.
2. Rinse
• Rinse handpiece with distal end pointing down under running cold
deionized, distilled or purified water for a minimum of 2 minutes.
• Use a sponge, soft lint-free cloth or soft bristle brush to assist in
removing gross soil.
Caution: Do not immerse. Do not use high pressure water or air. Do not
allow running water to enter distal end of handpiece or console connector.
3. Clean
• Prepare cleaning solution using alkaline detergent or neutral pH
enzymatic detergent per manufacturer's directions for correct
temperature, water quality (i.e. pH, hardness) and product
concentration/dilution.
• With distal end pointing down spray and wipe using cleaning
solution for a minimum of 2 minutes.
• Use a sponge, soft lint-free cloth or soft bristle brush to remove all
visible soil.
• Hold handpiece with distal end pointing down and insert ACB wetted
with cleaning solution into distal end of handpiece.
Warning: ACB is single use and should only be used to clean one
handpiece and associated attachments.
Caution: Do not immerse. Do not allow running water to enter distal end
of handpiece or console connector. Do not insert ACB beyond last bristle.
Do not force ACB into handpiece. Do not use pipe cleaners.
4. Rinse Thoroughly
• With distal end pointing down rinse device thoroughly using running
hot deionized, distilled or purified water for a minimum of 2 minutes.
• Hold handpiece with distal end pointing down and flush inside of
distal end using a syringe or pipette filled with hot deionized, distilled
or purified water.
Caution: Do not use high pressure water or air. Do not immerse. Do not
allow running water to enter distal end of handpiece or console connector.
5. Inspect
• Visually inspect handpiece for visible soil. Repeat Steps 2 thru 5 if
visible soil remains.
6. Go to Mechanical/Automated Cleaning section for further instructions.
8