4.0 INSPECTION
4.1 Acceptance And Transfer
1. Upon receipt of your Mizuho OSI 6875 hana
carton by following the provided instructions.
packaging. Visually inspect all surfaces for freight damage. Check each caster for proper
rolling operation.
NOTE: Any freight damage must be reported to the freight carrier immediately upon delivery.
It is the responsibility of the recipient to make freight damage claims.
2. Read the model/serial number identification label found at the head end of the table to
confirm the serial number and the input power requirements.
3. Place the 6875 hana
sides.
4. Perform Function Check; see section 5.0.
4.2 Inspection And Transfer
Before use, inspect the device for possible damage, excessive wear or non-functioning parts.
Visually inspect all accessible areas, electrical cords and all movable parts for possible
damage that may adversely affect the proper operation of the hana
Damaged or defective products should not be used or processed. Contact your local Mizuho
OSI sales representative for repair or replacement.
4.3 Pre-Procedure/Post-Procedure
Inspect and test the table as described in Function Check section 5.0.
Thoroughly clean the table as described in the Cleaning and Maintenance section 7.0.
Pay special attention to the cleanliness of the controls as excessive soil can affect
function.
Inspect the power cord for cuts in the insulation or damage to the connector.
On a smooth surface with the caster locks engaged, push the table. The table should not
move.
4.4 Semi-Annual
It is recommended to perform a Preventative Maintenance (PM) check on your 6875 hana
Surgery Table (see section 7.0 Cleaning and Maintenance).
A PM checklist is available from Mizuho OSI Customer Resource Group (800-777-4674).
MIZUHOSI 2009
®
Surgery Table in an area with at least 4 feet of clearance on all
®
Surgery Table, remove it from the shipping
Remove any protective wrapping or
15
®
Surgical Table.
®
NW0508 Rev. D