How supplied:
Mizuho OSI's femoral hook(s) and femoral hook support are supplied non-sterile. Cleaning
and sterilizing of the instrument(s) is required before each use according to your hospital's
washing, decontamination and sterilization procedures.
Contraindications:
The use of these surgical instruments is contraindicated when, in the judgment of the
physician, their use would be contrary to the best interest of the patient.
Precautions:
The Mizuho OSI's femoral hook(s) and femoral hook support are surgical instruments that
require special handling to prevent damage. Misuse can cause excessive stress or strain
that can result in damage that can adversely affect their intended use. Use caution during
cleaning and sterilization. The femoral hook(s) are approved for use only in combination with
the femoral hook support in conjunction with the Mizuho OSI 6850 PROfx
®
OSI 6875 hana
Surgery Table .
Care and handling: Inspection
Before each use, inspect the femoral hook(s) and femoral hook support for damage, wear
and functionality (check for nicks, burrs or bent parts).
Damaged or nonfunctioning instruments should not be used or processed. Contact your local
Mizuho OSI sales representative or Mizuho OSI Customer Resource Group for repair or
replacement.
Use of damaged instruments may increase the risk of tissue trauma, infection and
length of operative procedures.
Cleaning, disinfection, preparation for sterilization:
 Prior to each patient use inspect, clean, disinfect and sterilize the femoral hook(s) and
femoral hook support per hospital protocol for reprocessing surgical instruments.
 Do not use steel wool, wire brushes, pipe cleaners or abrasive detergents.
anything other than high quality brushes designed for surgical instrument cleaning may
result in damage.
 Disassembly of the femoral hook support before cleaning is generally not necessary
unless severely soiled or dictated by hospital policy.
 Do not use high acid (pH 4 or lower) or high alkaline (pH 10 or higher) products for
disinfectants, such as bleach and bi-chloride of mercury.