8.1. Alarms

When the concentrator sounds an alarm, a corresponding message will be displayed on the
screen. Take appropriate action as directed in the charts below.
8.1.1. Critical High Priority Alarms
IMPORTANT: These alarms will disable the device immediately.
Alarm message
EN
*Value: 01-20
8.1.2. High Priority Alarms
IMPORTANT: These alarms will allow the device to continue operating.
Alarm message
*Value: 21-50
8.1.3. Medium Priority Alarms
Alarm message
*Value: 51-70
8.1.4. Low Priority Alarms
Alarm message
*Value: 71-99
Description
Battery needs charging.
Battery is not an approved
battery
Service required.
Description
Device is unable to maintain
oxygen purity.
Estimated battery life less
than 17 minutes.
Service required.
Description
No breath detected for 15
seconds
Continuous flow of oxygen is
below specifications.
Service required.
Description
Service required.
14/185
Action
Recharge the battery pack by plugging in
to the power supply. Ensure all connections
are made securely.
Replace battery with an approved battery.
Contact your distributor.
Action
Be sure air inlet/outlet has not been
blocked. If alarm persists, contact your
distributor.
Charge the battery pack by plugging in to
power supply.
Important: The message will be automatical-
ly cleared when plugged in to power supply.
Contact your distributor.
Action
Check the cannula connection. Be sure to
breathe through nose, If alarm persists,
contact your distributor.
IMPORTANT: The message will be automati-
cally cleared when breathing is detected.
Check that cannula is not kinked and that
patient filter is properly installed. If alarm
persists, contact your distributor.
Contact your distributor.
Action
Contact your distributor.