Electromagnetic Compatibility (Emc) Information; Guidance And Manufacturer's Declaration: Electromagnetic Emissions - GCE Zen-O RS-00500 Manual Del Usuario

Ocultar thumbs Ver también para Zen-O RS-00500:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 94
Pulse mode bolus size (ml/breath) versus setting and breath rate
Breath per minute
15
20
25
30
35
40
All values +/- 15% over all operating conditions
Continuous Mode Flow (l/min) versus setting
Setting
0.5
1.0
1.5
2.0
All values +/- 0.2 litres over all operating conditions

14.1. Electromagnetic Compatibility (EMC) Information

Medical electrical equipment requires special cautions regarding electromagnetic compat-
ibility (EMC). Portable and mobile radio frequency (RF) communications equipment can affect
devices such as the Zen-O™ Portable Oxygen Concentrator. As such, the device should not be
used adjacent to other equipment. If this is not practical, then observe the device to make sure
it is operating properly at all times.

14.1.1. Guidance and manufacturer's declaration: electromagnetic emissions

The Zen-O™ Portable Oxygen Concentrator is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the concentrator should assure that it is used
in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
1
2
11
22
11
22
11
22
11
22
11
22
11
22
Compliance
Electromagnetic environment/guidance
Group 1
The Zen-O™ Portable Oxygen Concentrator uses RF en-
ergy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
Class B
The concentrator is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
Class A
network that supplies buildings used for domestic
purposes.
Complies
19/185
Setting
3
4
33
44
33
44
33
44
33
44
33
44
33
44
Flow rate
0.5
1.0
1.5
2.0
5
6
55
66
55
66
55
66
55
66
EN
55
57
50
50