Technical Description
14 .
Size:
Unit weight:
Power requirements:
EN
Purity:
Setting:
Inspiratory trigger
sensitivity:
Setting indicator:
Maximum oxygen
discharge pressure:
Humidity range:
Operating altitude:
Sound pressure level:
Type of protection
(electrical):
Degree of protection
(electrical):
Degree of protection
(water):
Degree of safety
(flammable anaesthetic
mixture):
Technical Description (continue)
Operating
temperature:
Storage temperature:
Alarm sound pressure
range:
Alarm system delays:
Oxygen concentrator
status indicator:
212 mm (W), 168 mm (D), 313 mm (H)
8.3" (W), 6.6" (D), 12.3" (H)
4.66 kg (10.25 lbs) (without carry bag and cart)
AC adaptor: 100-240V AC (+/- 10%), 50-60 Hz in, 24V DC, 6.25A out.
DC adaptor: 11.5-16V DC in, 19V, 7.9A out
(Important: see accessories list for model and part number of AC
power supply.)
87% - 96% at all flow rates, over operating conditions
User adjustable in 0.5 increments from 1.0 to 6.0 in pulse mode and
from 0.5 to 2.0 in continuous mode.
-0.12 cm/H
O
2
LCD display
20.5 psi
5% to 93% ± 2% non-condensing
0' to 9000' relative to sea level (0 km to 2.7 km), 1060 down to 700
mbar
42 dB(A) at setting 2 in pulse mode, tested according to ISO 3744
38 dB(A) at setting 2 in pulse mode, tested according to Prüfmeth-
ode 14-1 03/2007 MDS-Hi
Class II
Type BF
IP22 in carry bag (protection against small objects and tilted drip-
ping water)
IP20 out of carry bag (protection against small objects and no pro-
tection against water entering the concentrator)
Not suitable for use in the presence of a flammable anaesthetic
mixture
Continuous operation at temperatures between 5°C (41°F) and
40°C (104°F).
Between -20°C (-4°F) and 60°C (140°F).
65 to 85 dB(A)
Less than 10 seconds after detection (low oxygen alarms if oxy-
gen is less than 82% volume fraction at specified environmental
conditions)
High priority alarm that indicates when oxygen concentration
drops below 82%
18/185