Manual Cleaning

EN-8
1 Soak
• Prepare a fresh solution of detergent with
utility water according to the detergent
manufacturer's instructions.
• Immerse the device in the detergent solution,
ensuring the solution contacts all inner and
outer surfaces.
• Soak the device in the solution according to the
detergent manufacturer's recommendations.
2 Brush
• With the device immersed in the solution,
thoroughly brush the exterior with a soft-
bristled brush, focusing on dead-end holes,
mated or rough surfaces.
• Actuate and brush any movable parts in all
extreme positions.
• With the device immersed in the solution,
thoroughly brush any dead-end holes or
hard-to reach areas of the device using a bottle
brush.
3 Rinse
• Rinse the device with critical water until all
detergent residue is removed.
• Once all detergent residue is removed,
continue to rinse for 10 seconds.
• Drain excess water from the device and dry it
using a clean, lint-free cloth or pressurized air.
• Visually inspect each device for cleanliness,
paying close attention to hard-to-reach areas. If
visible soil remains, repeat steps 1-3.