Inspection And Testing - Stryker 0233032105 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 96
Maintenance,
Inspection And
Testing
Preparation For
Sterilization
Sterilization
EN-10
Inspect the tray before each use. If a problem
relating to functionality or appearance is
observed or suspected—such as (but not
limited to) corrosion, discoloration, pitting
or protrusions, binding, excessive wear, or
inability to fit or close the lid—discontinue
use immediately and contact your Stryker
representative.
Inspect all components for cleanliness. If fluid
or tissue buildup is present, repeat the above
processing procedures.
Clean, inspect, and prepare the tray as
recommended in this Processing Guide.
Ensure that devices are disassembled for
sterilization, if applicable.
Verify devices are compatible with both the
sterilization method and cycle intended to
process the tray.
Double wrap the tray prior to sterilization.
Warnings:
Assembled instruments will not sterilize
completely. Disassemble all instruments
prior to placement in the tray for
sterilization.
Use only the sterilization cycles outlined
in this document. Using unspecified
sterilization cycles may damage the device
or result in incomplete sterilization.
Note: For necessary materials and equipment, see
Materials and Equipment table.
Perform one of the following sterilization
cycles.

Publicidad

Tabla de contenido
loading

Tabla de contenido