Automated Cleaning
And Optional Thermal
Disinfection
Pre-Wash
(Optional)
Main Wash
Thermal
(optional)
Dry Phase
1. Place the device in the automated washer on an
incline to facilitate drainage.
2. Operate the washer-disinfector cycle.
The following minimum and maximum washer
parameters have been validated for cleaning
efficacy and device functionality:
Recirculation
1 – 2 minutes
Enzyme
Wash
2 – 4 minutes
Rinse 1
1 – 2 minutes
1 – 5 minutes
Rinse
3 – 15 minutes
3. Drain excess water from the device and dry it
using a clean lint-free cloth or pressurized air.
4. Visually inspect each device for cleanliness,
paying close attention to hard-to-reach areas. If
visible soil remains, repeat steps 1 – 4.
Note: Thermal disinfection value A
utilized for compatible devices. Do not exceed the
time and temperature specified in the table above.
Temperature
Time
1 minute
(110 – 122 °F)
(110 – 151°F)
(110 – 151°F)
(194 – 199°F)
High (≤115°C
Cold water
43 – 50°C
43 – 66°C
43 – 66°C
90 – 93°C
or 239°F)
≥3000 may be
0
Detergent
type
N/A
Enzymatic
Neutral or
Alkaline
N/A
N/A
N/A
EN-9