Steam (Autoclave)
EN-12
Wrapping
Temperature
Exposure
Time
Dry Time
(minimum in
chamber)
Warning: Only devices marked AUTOCLAVE
are compatible with steam sterilization
methods. Using steam sterilization on
devices that do not bear this marking can
cause permanent device damage.
Note: The water used in the autoclave process
must meet standards for clean steam per AAMI
ST79 Annex-Steam Quality.
Pre-vacuum (dynamic air removal)
U.S.
Outside of
132°C
134 – 137°C
(270°F)
(273 – 279°F)
4 minutes
3 – 5 minutes
60 minutes
50 minutes
Refer to Video Devices Processing Guide
1
P46385 to confirm device compatibility with
the Prion cycle.
Caution: Longer cycles, such as those
recommended for control or elimination of
Transmissible Spongiform Encephalopathies,
may be utilized. However, a device exposed
to longer cycles should be expected to have
reduced functional life.
Warning: Drying time depends on several
variables including altitude, humidity, type
of wrap, preconditioning, size of chamber,
mass of load, and placement in the
chamber. Users must verify that drying time
set in their autoclave yields dry surgical
equipment.
Prion Cycle
U.S.
Double
134°C (273°F)
18 minutes
50 minutes
1