Litter Top Removal - Stryker 1037 Manual De Instrucciones

Ocultar thumbs Ver también para 1037:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 107
English

LITTER TOP REMOVAL

Tools Required:
• 1/2" Socket with Extension
• 3/8" Drive Ratchet
• Standard Screwdriver
• 3/8" Drive Torque Wrench (ft-lb)
• 7/16"-20 Bolt
Procedure:
1 .
Using the foot pedal, pump the litter top all the way up .
2 .
Remove the stretcher mattress .
3 .
Using a 1/2" socket and a 3/8" drive ratchet, remove the truss head screws that hold the jack support tubes to the
jack shafts . When reinstalling, torque the truss head screws to 23 ± 3 ft-lb .
4 .
Insert a fine-thread 7/16"-20 bolt far enough into the top of the jack supports to separate the litter top from the
jack shaft .
5 .
With the assistance of another person, lift the litter straight up to remove it from the jack shafts and set it aside .
6 .
Reverse steps to reinstall the litter .
7 .
Verify proper operation of the unit before returning it to service .
Return To Table of Contents
1-26
Service Information
1037-009-006 REV D
www.stryker.com

Publicidad

Tabla de contenido
loading

Tabla de contenido