Instructions for Use
TPS-compact Mounting
Failure to comply with these installation instructions could result in the
WARNING
pump detaching from the system in the event of a rotor failure, which
could cause property damage or serious injury or death.
The TPS-compact can be installed in vertical position only.
Fix the TPS-compact in a stable position, mounting the inlet flange of the
turbopump to the system counter-flange.
The turbopump with ConFlat and ISO K inlet flange must be fixed to the vacuum
chamber by means of the appropriate Agilent hardware.
For further details see the "Installation of ConFlat and ISO K" section.
Power supply cord: The correct cable for electrical wiring is a three wires
(Ph+N+Earth) cable. The wire section has to be at least 0.75 mm
The TPS-compact cannot be fixed by means of its base or through other flanges
CAUTION
that are not inlet.
The System must be installed in a way that allows an easy interruption of
WARNING
the line voltage.
Assure a free space all around the system at minimum of 20cm to allow proper
CAUTION
air circulation.
170
2
(AWG18).