Instructions for Use

Stopping the TPS-compact

To switch off the TPS-compact it is sufficient to move the external switch to OFF
position. The integrated controller immediately stops the pumping system.
Before carrying out any work on the system, disconnect it from the power
WARNING
supply, vent the pump with the vent valve, wait until the rotor is completely
stopped and wait until the surface temperature of the pump is below 50° C.

Emergency stop

To immediately stop the TPS-compact in an emergency condition it is necessary to
remove the supply cable from the mains plug or moving the external switch to OFF
position.
This mode is to be used only in an emergency. The use of this stop mode
WARNING
could lead to faults and/or damages at the pump. Improper use of this
stop mode could void the warranty.