Limitations

The ASTUTE140
Meter times the incubation period for the test, and only one test can be run at a time;
therefore, each NEPHROCHECK
loaded with the patient sample or liquid control. The time delay in testing that would
result from loading more than one cartridge at a time may lead to inaccurate test results.
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
Meter does not contain a "batch mode" function. The ASTUTE140
®
®
Test cartridge must be run immediately after being
Introduction
®