Error Messages
Type
*Error codes may begin with 001, 002, 003 or 004.
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
(Continued)
Screen Display
Maintenance and Troubleshooting
Error Code*
What it Means
Unknown
00X.061.031
Lot Error. The
required Lot is
not registered.
00X.061.008
Cartridge
Lot Locked.
Passing positive
and negative LQC
measurements
for the test lot is
required.
00X.061.046
Expired liquid
lot. The LQC lot
registered is past
the expiration
date.
Action to Take
Register a new test
lot (See "Test Lot
Registration" on
p. 4).
If the problem
persists, call Astute
Technical Support
(See "Contacting
Astute Medical, Inc."
on p. 7 for contact
information).
Perform LQC for
the Test lot. After
receiving passing
measurements for
both positive and
negative LQC, restart
the entire testing
procedure with a new
cartridge.
If the LQC fails
repeatedly or the
problem persists,
call Astute Technical
Support (See
"Contacting Astute
Medical, Inc." on
p. 7 for contact
information).
Passing positive
and negative LQC
measurements
for the test lot is
required.
Register and perform
testing with a new
LQC lot. A passing
measurement for
both positive and
negative LQC is
required prior to
running a patient
sample.