ASTUTE140
®
6. When the EQC procedure completes, the results will be displayed along with the
word Passed if the meter passed the EQC procedure or Failed if it did not. If the
EQC procedure failed, run the EQC procedure again following the steps below:
a. Press the right soft key to display the Options pop-up menu.
b. Use the
c. When the drawer automatically opens, remove the ASTUTE140
Device and repeat steps 4 and 5. NOTE: If a second ASTUTE140
Device is available, the second device should be used.
d. When the EQC procedure completes, the results will be displayed along with
the word Passed if the ASTUTE140
Failed if it did not. If the procedure failed, contact Astute Technical Support
(See "Contacting Astute Medical, Inc." on p. 7 for contact information).
7. To manually print the EQC results if Autoprint is disabled:
a. Press the right soft key to display the Options pop-up menu.
b. When the Options pop-up menu is displayed, Print is highlighted. To print
the EQC results, press the right soft key.
c. To hide the Options pop-up menu, use the
Options and press the right soft key.
d. To return to the Main Menu, press the left soft key.
8. After completing the EQC procedure, press the
screen to open the ASTUTE140
Device and store it in the case provided away from direct light.
Test Lot Registration
Assay-specific test cartridges that are uniquely compatible with the ASTUTE140
are used to test patient samples. The NEPHROCHECK
with the ASTUTE140
only NEPHROCHECK
be used with the meter. For a list of approved tests for use with the ASTUTE140
and to purchase assay-specific test cartridges, go to www.astutemedical.com or consult
your Astute Medical, Inc. sales representative.
Each new test lot must be registered prior to use of any NEPHROCHECK
from that lot. Operators may register the test lot only if the Supervisor granted registration
permission in the default settings; otherwise, Supervisors must register the test lot.
See the NEPHROCHECK
52
Meter Operation
key to highlight Next Control and press the right soft key.
®
®
Meter should not be used with any other testing device. Similarly,
Test cartridges designed for use with the ASTUTE140
®
Test Kit Package Insert for storage instructions.
®
Meter passed the EQC procedure or
®
key to highlight Hide
key to the left of the LCD
Meter drawer. Remove the ASTUTE140
®
Test cartridges designed for use
ASTUTE140
EQC
®
EQC
®
EQC
®
Meter
®
Meter can
®
Meter
®
®
Test cartridges
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
Meter
®