ASTUTE140
Meter
®
10. If LQC measurements for the NEPHROCHECK
a screen indicating the amount of time remaining until test completion will be
displayed. To abort the test for any reason, select Cancel by pressing the left soft
key.
11. Once the test is complete, the test results will be displayed.
12. To manually print the patient test results if Autoprint is disabled:
13. Press the eject button to the left of the LCD screen to open the ASTUTE140
Meter drawer. Remove the NEPHROCHECK
to the laboratory's standard practices for the type of biological sample used.
14. To run the next patient sample after viewing or printing the patient test results,
press the right soft key to display the Options pop-up menu. Use the
highlight Next Patient and press the right soft key. Repeat steps 4–13.
Review and Management of Test Results
Patient, LQC, EQC and Calibration Verification (Available in the United States only)
test results can be recalled, printed or sent to the LIS at any time. To perform these
functions, follow the instructions below.
Patient Results
1. Press the
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
iii. Run high and low LQC measurements (See "External Liquid Quality
Control" on p. 56 for instructions).
iv. When high and low LQC measurements have passed, repeat steps
1–9.
a. Press the right soft key to display the Options pop-up menu.
b. When the Options pop-up menu is displayed, Print is highlighted. To print
the patient test results, press the right soft key.
c. To hide the Options pop-up menu, use the
Options and press the right soft key.
d. To return to the Main Menu, press the left soft key.
key to display the Main Menu.
ASTUTE140
Meter Operation
®
Test cartridge lot have passed,
®
key to highlight Hide
Test cartridge and discard according
®
®
key to
65