To run the EQC procedure, perform the following steps using the ASTUTE140
Device provided with the ASTUTE140
1. Press the
successful log in, the Main Menu will automatically be displayed).
2. Use the navigation keys to highlight the Run EQC icon.
3. Press the right soft key to select Run EQC. The ASTUTE140
automatically open.
4. When prompted, place the ASTUTE140
the ASTUTE140
ASTUTE140
it clicks.
Figure 56.
Inserting ASTUTE140
®
EQC Device
5. After closing the drawer, a temporary screen will appear indicating that the
ASTUTE140
time remaining until test completion will be displayed.
Figure 57.
Temporary screen showing
EQC procedure countdown
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
Meter:
®
key to display the Main Menu (if running EQC immediately after
®
Meter drawer with the Astute Medical logo marked on the
EQC Device pointed toward the meter. Gently close the drawer until
®
Meter is running the EQC procedure and displaying the amount of
®
ASTUTE140
®
Meter drawer will
®
EQC Device into the recess inside
®
Meter Operation
EQC
®
51