External Liquid Quality Control (Lqc) - Astute Medical ASTUTE140 METER Manual Del Usuario

Ocultar thumbs Ver también para ASTUTE140 METER:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ASTUTE140
®
9. When prompted, hold the NEPHROCHECK
liquid control to be registered next to or against the numeric keypad to register
the liquid control lot information and press the right soft key to select OK.
10. If registered correctly, a screen indicating that the liquid control lot number was
successfully read from the NEPHROCHECK
and the lot number will be displayed. Press the right soft key to select Accept. The
lot that was just registered should now appear in the list of Registered LQC Lots.
11. If registered incorrectly, an error message will appear. Press the right soft key
to select OK and close the error message. Repeat steps 9–10. If registered
incorrectly a second time, contact Astute Technical Support (See "Contacting
Astute Medical, Inc." on p. 7 for contact information).
12. After use, return the NEPHROCHECK
store with the corresponding liquid control.
13. To register the second liquid control in the set, or to register another liquid control
set from a different lot, repeat steps 7–12.

External Liquid Quality Control (LQC)

High and low liquid controls are available to assist the Laboratory in monitoring system
performance, including the test reagents and the ASTUTE140
It is recommended that LQC be performed at the following times:
Every 30 days
P
With each new lot number of NEPHROCHECK
P
With each new shipment of NEPHROCHECK
P
After ASTUTE140
P
In accordance with the laboratory's standard quality control procedures
P
56
Meter Operation
Meter servicing
®
Liquid Control RFID Card for the
®
Liquid Control RFID Card will appear,
®
Liquid Control RFID Card to its sleeve and
®
Meter.
®
®
Test cartridges
Test Kits
®
ASTUTE140
Figure 61.
Register Liquid QC Lot
screen
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07