ASTUTE140
Meter
®
Error Messages
Type
*Error codes may begin with 001, 002, 003 or 004.
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
(Continued)
Screen Display
Maintenance and Troubleshooting
Error Code*
What it Means
The
00X.061.049
Passwords are
Different.
User Cannot
00X.061.047
Be Deleted.
The last or only
supervisor cannot
be deleted.
Barcode
00X.061.010
Transmission
Error. The bar
code was not
read correctly.
Action to Take
Carefully re-enter the
password and retry.
Use the exiting
supervisor or create
a new user.
Press OK and repeat.
If the error
persists, check
that the barcode
isn't damaged or
incomplete.
79