ASTUTE140
Meter
®
8. When prompted, place the prepared NEPHROCHECK
recess inside the ASTUTE140
on the NEPHROCHECK
drawer until it clicks.
Figure 64.
Running LQC
9. After closing the drawer, a temporary screen will appear indicating that the
ASTUTE140
remaining until test completion will be displayed.
10. When the LQC procedure completes, the results will be displayed along with the
word Passed if the procedure passed or Failed if it did not. If the LQC procedure
failed, run the LQC procedure again following the steps below:
11. Press the
drawer. Remove the cartridge and discard according to the laboratory's standard
practices for the type of biological sample used.
12. To manually print the LQC results if Autoprint is disabled:
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
®
Test cartridge pointed toward the meter. Gently close the
Meter is running the LQC procedure and the amount of time
®
a. Press the right soft key to display the Options pop-up menu.
b. Use the
key to highlight Next Control and press the right soft key.
c. Repeat steps 4–9 using a freshly prepared NEPHROCHECK
When the LQC procedure completes, the results will be displayed along with
the word Passed if the meter passed the LQC procedure or Failed if it did not.
If the procedure failed a second time, contact Astute Technical Support
(See "Contacting Astute Medical, Inc." on p. 7 for contact information).
key to the left of the LCD screen to open the ASTUTE140
a. Press the right soft key to display the Options pop-up menu.
ASTUTE140
Meter drawer with the Astute Medical logo marked
®
Meter Operation
®
Test cartridge into the
®
Test cartridge.
®
Meter
®
59