The required frequency for running LQC may be set by a Supervisor in the general
ASTUTE140
instructions).
Prior to running the first LQC procedure, both the high and low liquid controls in the
liquid control kit must be registered if they are from a new liquid control lot (See "Liquid
Control Lot Registration" on p. 54 for instructions).
The LQC procedure must be run twice: once with the high liquid control and once with
the low liquid control. For each lot of NEPHROCHECK
low liquid controls must have passed within the time window set in the default settings
before patient samples can be run using NEPHROCHECK
To run the LQC procedure, perform the following steps and repeat for the second
control in the liquid control set:
1. Press the
successful log in, the Main Menu will automatically be displayed).
2. Use the navigation keys to highlight the Run LQC icon.
Figure 62.
Main Menu screen with
Run LQC icon highlighted
3. Press the right soft key to select Run LQC.
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
Meter default settings (See "Quality Control Settings" on p. 31 for
®
key to display the Main Menu (if running LQC immediately after
ASTUTE140
Meter Operation
®
Test cartridges, both high and
®
®
Test cartridges from that lot.
57