Maintenance and Troubleshooting
Error Messages
Type
Screen Display
*Error codes may begin with 001, 002, 003 or 004.
86
(Continued)
Error Code*
00X.003.001
00X.061.055
00X.061.014
What it Means
Action to Take
System
If you are able, turn
Failure. A
the Meter off and
system failure
then back on. Run
occurred during
an EQC. If it passes,
the last test.
repeat the whole test
procedure for the
last test performed.
If the EQC fails or
if this happens
repeatedly, call
Astute Technical
Support. (See
"Contacting Astute
Medical, Inc." on
p. 7 for contact
information).
Wrong entry.
Review and modify
Wrong format
the format in which
or date range
the information was
entered.
entered.
Lot Data
Delete existing
Base Full.
entries in order to
The database
clear memory and
containing the
register new lots.
Lots previously
registered is full.
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
ASTUTE140
Meter
®