3. When the Run Patient screen is displayed, Patient ID will be highlighted. When
prompted for Patient ID we suggest entering 001, 002, 003, 004, 005, (or unique
alphanumeric codes as defined by your Laboratory) corresponding to each
NEPHROCHECK
using the numeric keypad or an external keyboard (if connected). NOTE: The
NEPHROCHECK
the calibration verification sample.
4. Using the grips on the side of the NEPHROCHECK
the cartridge inside the ASTUTE140
logo towards the inside of the meter drawer. Keep the NEPHROCHECK
cartridge horizontal and avoid tipping the test cartridge during placement into the
ASTUTE140
5. Close the ASTUTE140
and Verification result will be displayed.
6. Once the calibration verification test is complete, the test results will be displayed.
7. To manually print individual calibration verification test results if Autoprint is
disabled:
8. Press the eject button to the left of the LCD screen to open the ASTUTE140
Meter drawer. Remove the NEPHROCHECK
cartridge, the calibration verification vial, and the conjugate vial in accordance
with local regulations.
9. Press the right soft key to display the Options pop-up menu. Use the
highlight Next Patient and press the right soft key. Repeat steps for the remaining
four levels of NEPHROCHECK
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
®
Calibration Verification Vial level. Enter the Patient ID manually
Test cartridge must be run immediately after inoculation with
®
Meter drawer.
®
Meter drawer. In approximately 20 minutes the Calibration
®
a. Press the right soft key to display the Options pop-up menu.
b. When the Options pop-up menu is displayed, Print is highlighted. To print
the individual calibration verification test results, press the right soft key.
c. To hide the Options pop-up menu, use the
Options and press the right soft key.
d. To return to the Main Menu, press the left soft key.
ASTUTE140
Test cartridge, position
®
Meter drawer with the Astute Medical
®
key to highlight Hide
Test cartridge and discard the
®
Calibration Verification Kit Vials.
®
Meter Operation
®
®
Test
®
key to
61