Maintenance and Troubleshooting
Error Messages
Type
Screen Display
*Error codes may begin with 001, 002, 003 or 004.
80
(Continued)
Error Code*
00X.061.042
00X.061.044
00X.061.043
What it Means
Action to Take
Cartridge
Press eject button
Error. The bar
to open the drawer.
code on the
Without removing the
cartridge is
cartridge or touching
invalid, missing
the window, check
that the barcode
or unreadable.
on the cartridge
is not damaged or
incomplete. Close
the drawer and retry.
If the error reoccurs,
retest with a new
Cartridge with a valid
barcode.
If the problem
persists, see Astute
Medical Technical
Support. (See
Contacting Astute
Medical, Inc." on
p. 7 for contact
information).
Device Type
Check the type of
Error. The device
cartridge you are
type is incorrect
testing and in which
for the testing
screen you are
being conducted.
operating in from
the Main Menu (i.e.,
running a patient
sample or EQC
device). Retest using
a Cartridge of the
correct device type.
Expired
Properly discard
Cartridge. The
the expired test lot
Cartridge is past
and use a valid test
the expiration
lot. Always check
date.
the expiration date
printed on the kit box
prior to beginning a
test procedure.