Installation
6. When the QC Settings screen is displayed, LQC Interval (Days); 0 = off will be
highlighted. Use the numeric keypad, or an external keyboard (if connected) to
enter the interval in days at which LQC must be run. NOTE: entering "0" indicates
that LQC is not required to be run at all. To change an existing entry, use the
key to first delete the entry, then enter the new number using the numeric keypad
or an external keyboard (if connected).
7. Press the
keypad, or an external keyboard (if connected) to enter the interval in hours at
which EQC must be run. NOTE: entering "0" indicates that EQC is not required to
be run at all. To change an existing entry, use the
then enter the new number using the numeric keypad or an external keyboard (if
connected).
8. Press the right soft key to accept the intervals entered.

LIS Settings

If the ASTUTE140
interval, the IP Address, Port Number, and whether results will be automatically
transmitted from the meter to the LIS. To set the LIS settings, follow these steps:
1. Press the
2. Use the navigation keys to highlight the Supervisor Menu icon.
3. Press the right soft key to display the Supervisor Menu.
4. When the Supervisor Menu is displayed, General Meter Settings is highlighted.
Press the right soft key to display the General Settings screen.
5. When the General Settings screen is displayed, Default Settings is highlighted.
Use the
32
key to highlight EQC Interval (Hours); 0 = off. Use the numeric
®
Meter is connected to an LIS, Supervisors can set the timeout
key to display the Main Menu.
key to highlight LIS Settings and press the right soft key.
key to first delete the entry,
Figure 26.
Main Menu screen with
Supervisor Menu icon
highlighted
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
ASTUTE140