Inspection, maintenance and checks
►
Allow the product to cool down to room temperature.
Note
Aesculap additionally recommends the occasional spraying of moving parts
(such as knobs, connections) with STERILIT Power Systems oil spray.
►
Inspect the product after every cleaning and disinfecting cycle for:
cleanliness, damage, function, abnormal operation noise, excessive
heat or strong vibration.
►
Inspect the saw blades for broken, damaged or blunt cutting edges.
►
Set aside the product if it is damaged.
7.10 Packaging
►
Observe the instructions for use of the respective packaging and hold-
ers (e.g. Instructions for use TA009721 for Aesculap ECCOS holder sys-
tem).
►
Insert product 1 in the correct position in the ECCOS holders, see Fig. B.
►
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
►
Ensure that the packaging will prevent a recontamination of the prod-
uct.
7.11 Steam sterilization
CAUTION
Damage or destruction of the batteries due to processing!
►
Do not sterilize the rechargeable battery.
Note
Remove all attached components from the product (tools, accessories)
before sterilization.
►
Check to ensure that the sterilizing agent will come into contact with
all external and internal surfaces (e.g. by opening any valves and fau-
cets).
►
Use a validated sterilization method:
– Steam sterilization in fractionated vacuum process
– Steam sterilizer in accordance with DIN EN 285 and validated in
accordance with DIN EN ISO 17665
– Sterilization in fractionated vacuum process at 134°C, holding time
5 min
When sterilizing multiple products in one steam sterilizer:
►
Ensure that the maximum permissible load of the steam sterilizer
according to the manufacturer's specifications is not exceeded.
7.12 Storage
►
Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
10
8.
Maintenance
To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.