Inspection, maintenance and checks

Allow the product to cool down to room temperature.
Note
Aesculap additionally recommends the occasional spraying of moving parts
(such as knobs, connections) with STERILIT Power Systems oil spray.
Inspect the product after every cleaning and disinfecting cycle for:
cleanliness, damage, function, abnormal operation noise, excessive
heat or strong vibration.
Inspect the saw blades for broken, damaged or blunt cutting edges.
Set aside the product if it is damaged.

7.10 Packaging

Observe the instructions for use of the respective packaging and hold-
ers (e.g. Instructions for use TA009721 for Aesculap ECCOS holder sys-
tem).
Insert product 1 in the correct position in the ECCOS holders, see Fig. B.
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
Ensure that the packaging will prevent a recontamination of the prod-
uct.

7.11 Steam sterilization

CAUTION
Damage or destruction of the batteries due to processing!
Do not sterilize the rechargeable battery.
Note
Remove all attached components from the product (tools, accessories)
before sterilization.
Check to ensure that the sterilizing agent will come into contact with
all external and internal surfaces (e.g. by opening any valves and fau-
cets).
Use a validated sterilization method:
– Steam sterilization in fractionated vacuum process
– Steam sterilizer in accordance with DIN EN 285 and validated in
accordance with DIN EN ISO 17665
– Sterilization in fractionated vacuum process at 134°C, holding time
5 min
When sterilizing multiple products in one steam sterilizer:
Ensure that the maximum permissible load of the steam sterilizer
according to the manufacturer's specifications is not exceeded.

7.12 Storage

Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
10
8.

Maintenance

To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.