Troubleshooting list
►
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Fault
Cause
Product not running
No battery
Battery not charged
Battery broken
Product in secured OFF position
Product defective
Product becomes too hot Excessive use
Reprocessing/maintenance carried
out incorrectly
Fall damage, product defective
Blunt tool
Insufficient power
Product broken
Blunt tool
Loud running noise
Broken gearing/ball bearings of the
product
Lid cannot be
Lid not compatible
attached/removed
Lid deformed/broken
Connection on product
deformed/broken
Battery cannot be
Battery not compatible
inserted/removed
Battery deformed/broken
Battery shaft to product
deformed/defective
Sterile funnel or battery
Sterile funnel or battery removal
removal device cannot be
device not compatible
attached
Sterile funnel or battery removal
device deformed/broken
Battery compartment at product
deformed/broken
Detection
No battery in the battery shaft
No signal tone when inserting
the battery
No signal tone when inserting
the battery
Safety catch is in OFF position
Product not running
Product heating
Product heating
Product heating
Tool and product become hot
Insufficient product perfor-
mance
Steep rise in temperature after a
short period of time
Tool blades worn
Loud, perceptible noise during
operation
Lid does not click into place
Lid is difficult or cannot be
mounted/removed
Lid is hard or impossible to
install/remove
Battery cannot be completely
inserted into the battery com-
partment
Battery is hard or impossible to
insert/remove
Battery is difficult or cannot be
mounted/removed
Sterile funnel or battery removal
device not attachable to the bat-
tery shaft
Sterile funnel or battery removal
device difficult or not attachable
Sterile funnel or battery removal
device is hard or impossible to
attach
Remedy
Insert battery.
Charge battery in charger.
Have the manufacturer repair the battery.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Observe instructions for use (nominal operating
mode).
Observe instructions for use (processing, care).
Preventive measure: Oil product before every ster-
ilization.
Have the manufacturer repair the product.
Change tool.
Observe instructions for use (processing, care).
Preventive measure: Oil product before every ster-
ilization.
Have the manufacturer repair the product.
Observe instructions for use (nominal operating
mode).
Have the manufacturer repair the product
Change tool.
Follow operating instructions (preparation, care).
Preventative: lubricate product before every steril-
ization.
Have the manufacturer repair the product
Use the compatible lid for GA334.
Have the manufacturer repair the lid.
Have the manufacturer repair the product.
Use the compatible battery for GA334.
Have the manufacturer repair the battery.
Have the manufacturer repair the product
Use suitable sterile funnel or battery removal
device for GA334.
Change sterile funnel or battery device.
Have the manufacturer repair the product.
en