Fault
Cause
Speed control knob can-
Product in locked OFF position
not be actuated
Speed control knob sticks/is defec-
tive
Saw blade does not move Noisy motor
Insufficient cutting per-
Saw blade teeth worn
formance of the saw
Battery power is low
blade
Battery empty/broken Symbol "Bat-
tery replacement recommended"
illuminated on the charger
The saw blade cannot be
Debris on the saw blade or in the
coupled
coupling
Deformed coupling parts/saw blades Deformation
-
Saw blade cannot be
Saw blade can be pulled out without
locked and can be pulled
rotating sleeve action
out during function check
Saw blade cannot be
-
uncoupled
Saw blade coupling warped or bent
saw blade
Support foot cannot be
Support foot cannot be coupled
coupled
Support foot cannot be coupled
Marking on support foot points to
"Closed" symbol.

10. Technical Service

DANGER
Danger to life of patients and users in case of malfunctions and/or
failure of protective measures!
Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.
CAUTION
Modifications of medical devices may result in a loss of potential
guarantee/warranty claims and forfeiture of applicable licenses.
Do not modify the product.
Contact national B. Braun/Aesculap representative for service and
repair.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the address indicated above.
12
Detection
Safety catch is in OFF position
Speed control knob cannot be
actuated
Broken gearings
Blunt saw blade
Power/speed of drive inadequate Charge battery
Debris
Incompatible saw blade
Rotating sleeve in incorrect
position (not returning automat-
ically to rest position)
Rotating sleeve in incorrect
position
Mechanical damage
Saw blade bent
Support foot bent
Locking sleeve is closed
Remedy
Turn the safety catch to the ON position.
Have the manufacturer repair the product.
Have product repaired by the manufacturer
Replace saw blade
Replace battery
Clean saw blade/coupling with cleaning
template 7
Have product repaired by the manufacturer
Insert a compatible Aesculap saw blade, see Acces-
sories/Spare parts
Clean clutch if soiled
If damaged, have device repaired by the manufac-
turer
Turn rotating sleeve 180° in direction of the arrow,
see Attaching and detaching the saw blade
Have repaired by manufacturer
Replace saw blade
Have support foot repaired by the manufacturer
Open locking sleeve, see Coupling the support foot.