Applicable standards
EN
Essential performance
requirements
Measurement accuracy
2. Technical data relating to the power supply
Specifications
Dimensions (W x H x D)
Weight
Shelf life
Type
Nominal capacity
Nominal voltage
Charging voltage
Charge time (0% to 95%)
Recommended charging
temperature
Certifications
46
EN 60601-1:2006/A1:2013/A12:2014
EN 60601-1-2:2015/A1:2021
EN 60601-1-12:2015
EN 62366-1:2015
EN 62304:2006/A1:2015
ISO 18562-1:2017
ISO 18562-2:2017
ISO 18562-3:2017
Under normal conditions of use, EOlife® must display the bar graph in real time The
screen must not freeze
Volume measurements are based on FlowSense® flow sensor measurements and are
expressed in mL for the BTPS (body temperature and pressure, saturated)
The measurement accuracies of the values displayed on the screen are as follows:
Vi (volume insufflated): ± 8.8% of the actual value measured under normal
conditions of use
Vt (tidal volume) without leakage: ± 8.8% of the actual value measured under
normal conditions of use
Freq (ventilation frequency): ± 1 cycle per minute
FlowSense® data:
Flow range: ± 250 slm (standard litre per minute)
Dead space: < 10 ml
Update time: 0 5 ms
Note: Certain types of ventilation bag can affect the measurement accuracy due to
their design (non-laminar outgoing air flow). A slight measurement deviation may be
seen but which has no impact on compliance with regulatory requirements
73 mm x 40 mm x 14 mm
50 g ± 5 g
6 months (at a storage temperature between -10°C and +35°C)
Warning:
Battery life may be reduced if it is stored outside this temperature range. Battery
life can be extended by charging the battery.
Li-ion
1,900 mAh (7 03 Wh)
3 7 V
12-24V CV
Approx 2 hours
+10°C to 45°C
IEC 62133
UN38 3
UL1642
EOlife® battery