Össur Logic App
For the initial fitting, the healthcare professional must use the Össur Logic app. It is available on Apple App Store.
See Össur Logic on the App Store for mobile device compatibility.
The patient can use the Össur Logic app to connect to the device and access some functions.
DEVICE SELECTION
See Össur catalog for a list of recommended components.
Use Össur titanium adapters to achieve waterproof performance.
Note: Adapter installation and set-screw tightening shall be performed according to instructions provided in their
respective accompanying documents.
Verify that selected variant of the device is suitable for the impact level and weight limit according to the following
table.
Warning: Do not exceed weight limit. Risk of device failure.
Incorrect category selection may also result in poor device function.
Table 1
Weight kg
Weight lbs
Low Impact Level
Moderate Impact Level
ASSEMBLY INSTRUCTIONS
Foot Cover and Spectra Sock
Caution: To avoid pinching fingers, always use a shoehorn.
1. Put the foot into the Spectra Sock.
2. Remove the foot cover attachment.
3. Use the applicable (straight) end of a shoehorn to put the foot with the Spectra Sock into the Foot Cover.
4. Move the shoehorn up to fully push the foot into the Foot Cover.
5. Put the foot cover attachment back onto the Foot Cover.
6. After alignment is complete, fix the Spectra Sock to the prosthesis to seal against dust and dirt.
Note: The Spectra Sock must be pulled up to prevent it from interfering with moving parts of the foot.
If required doff the Foot Cover as follows:
1. Remove the foot cover attachment.
2. Insert the applicable (edged) end of a shoehorn behind the foot.
3. Push the shoehorn down and pull the foot out of the Foot Cover.
4. Fully remove the Spectra Sock.
Prosthesis
Assemble prosthesis with applicable devices.
Warning: Risk of structural failure. Components from other manufacturers have not been tested and may cause
excessive load on the device.
Warning: Ensure proper attachment by following the applicable device assembly instructions.
Caution: Do not use heel wedges. Heel wedges will cause premature wear and tear.
ALIGNMENT INSTRUCTIONS
Before alignment and setup the device should be powered ON and connected to the Össur Logic App.
Initial Setup
Turn the device on.
After turning the device ON, two orange lights on the LED Indicator Ring will show initial set-up is required.
The setup consists of five steps.
1. Ankle Alignment
2. Bench Alignment
45-52
53-59
60-68
99-115
116-130
131-150
1
1
1
2
69-77
78-88
151-170
171-194
2
3
4
3
4
5
89-100
101-116
117-125
195-220
221-256
257-275
5
6
6
7
7
8
5