Descargar Imprimir esta página

Össur PROPRIO FOOT PSX02 Instrucciones Para El Uso página 9

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 46
Charging
1. Plug the power supply into the wall socket.
2. Plug the power supply into the charging port on the back of the device.
3. After a short while, the battery indicator will show the battery charge level with solid green lights and indicate
charging with a flashing green light. If the power supply is disconnected during charging, the battery indicator
may continue flashing for a short while.
4. After charging, disconnect the power supply from the charging port.
5. Disconnect the power supply from the wall socket to save energy. 
Device will turn ON automatically when the power supply is disconnected.
Note: It takes approximately 4 hours to charge the device battery pack when fully depleted.
Charge the battery daily for optimal performance.
Note: The battery pack cannot be overcharged.
Note: A fully charged battery pack can last for 24 to 48 hours, depending on activity level.
Note: The battery pack can only be charged in temperatures between 0°C and 45°C. If charge is attempted outside
these limits, it will be delayed until the battery pack temperature is within the specified range.
Battery Charge Level Indication
Push the battery indicator button. The green lights on the battery indicator show approximate battery level.
0%
Charge Level
- 15%
Indicator
appearance
Ankle Alignment (while wearing prosthesis)
Ankle Alignment adjusts the angle of the ankle to allow optimal posture with shoes of various heel heights.
1. Patient should be in a seated position.
2. Ensure that the foot is positioned on a flat surface, with the sole from heel to toe on the ground.
3. Initiate Ankle Alignment by pressing the power button quickly twice successively or in the app.
4. The device will beep once confirming initiation, and the indicator will light up, showing the alignment progress
(Table 2).
5. When device beeps again, lift the foot from the ground to allow for ankle adaptation.
6. The actuator will move the ankle into appropriate position.
7. If successful, the device will emit an ascending audio tone, and the indicator will flash white. If unsuccessful,
the indicator will flash yellow, showing process needs to be repeated.
Supported activities
If device is unable to move to the desired position, it will stop moving.
Note: The device can only move while unloaded in swing phase. Short swing phase duration (under 0.4 sec), heavy
shoes, interrupted swing cycle, or foot contact with the ground can result in incomplete dorsiflexion movement.
Walking on Level Ground
• The device automatically adapts to the patients walking speed and style.
Note: Circumduction or any abnormal side movement may hinder the initiation of dorsiflexion during swing phase.
Inclined Ascent
• On inclines, the device gradually raises the toe to the degree of the slope.
• During swing phase, the device will raise the toe to provide additional ground clearance.
Inclined Descent
• On declines, the device lowers the toe.
Note: It is not necessary to lead with a particular limb when ascending or descending slopes.
Warning: Ascending or descending inclines of more than 15° may trigger stair response.
16%
30%
40%
- 29%
- 39%
- 49%
50%
60%
70%
- 59%
- 69%
- 79%
80%
90%
95%
- 89%
- 94%
- 100%
9