MULTIJET PRO
10.0 - FILTER REMOVAL AND CLEANING
The Multijet PRO has filters inside the test tubes to retain residues that can cause malfunctioning of the drain
valves.
To clean the filters, ensure that the Multijet is off.
Demountig
(see figures)
1 - Remove the upper mounting rings.
2 - With the test tubes free, gently slide the top of each one until it is outside of the structure. Lift out the set.
3 - Remove the rubber base.
4 - For cleaning, remove the filters from the rubber base. If necessary, the valve can also be cleaned inside.
5 - Remove the plunger (use magnetized tools).
6 - Gently clean the internal parts, use cotton swab or similar.
Do not use sharp objects, sharp or that could damage the rubber. e.g.: screwdriver.
IMPORTANT: Protect the compartments of the plungers, do not let that be invaded by foreign objects,
or insects.
CAUTION: The tubes are made of glass. During the procedure keep them free of impacts.
Demounting
2
3
1
6
4
5
39