14. Place the IPG in the subcutaneous pocket with the etched writing "This Side Up" facing the skin
and parallel to the skin surface.
Note: Make the pocket no deeper than 2 cm for rechargeable IPGs and no deeper than
2.5 cm for Vercise Genus non‑rechargeable IPGs. Communication, including device
programming could become ineffective at depths greater than 2.5 cm. For rechargeable
IPGs, IPG charging could become ineffective at depths shallower than 0.5 cm or greater
than 2 cm.
Incorrect placement of the IPG in the pocket could require a revision surgery.
Warning:
a. Coil the excess Lead Extension length under or around the IPG perimeter.
Avoid placing the excess length of the Lead Extensions on the superficial
Warning:
surface of the IPG as this may result in tissue erosion, ineffective
communication, or charging difficulty.
b. Optional: Secure the IPG to the fascia by suturing through the holes in the IPG Header.
15. Close the incisions.
Caution: Be careful not to damage the DBS Lead, IPG, or other implanted components
when closing the incisions.
Note: When closing the incision over the Lead Extension Connector, orient the Lead
Extension Connector to minimize the profile under the skin.
Implanting the DBS System
Surgical Implant Manual
92495783-02