Vercise™ Deep Brain Stimulation Systems
Connecting to the IPG
When connecting the Lead Extension to the IPG, it is recommended that the following connection
configurations are utilized.
2‑Port (16 Contact) IPG
For dual 8 Contact Lead connection with the bifurcated Lead Extension DB‑3128 to the
2‑Port IPG:
7
a. Connect the left hemisphere Extension tail with the gold marker band to port L (or C)
(Figure 35).
b. Connect the right hemisphere Extension tail with no marker band to port R (or D)
(Figure 36).
7 DB-3128 is a 2x8 Extension and can connect up to two (2) 8 Contact DBS Leads to a 2-Port/16 Contact IPG.
Surgical Implant Manual
92495783-02
Page 46 of 1157
L
Figure 35. Connection to IPG Port L
Figure 36. Connection to IPG Port R
R