USER INTERFACE SCREENS
Mean ICP Histogram Screen
The ICP Histogram screen shows the distribution of mean ICP values over a period of time. Additionally, the Time Above Threshold
is calculated. Time above threshold is the total time the mean ICP has been above one of two user-defined thresholds within the
selected timeframe (refer to Analysis Threshold Setup Screen section for information on how to set the thresholds).
NOTE: The same thresholds used for the PTD calculations in the Mean ICP Trend Screen are used for the Time Above Threshold
calculation.
A
B
C
D
A
ICP Time Above Threshold
1. T = Time above threshold (hours, minutes and %).
2. (xx) = Pressure Threshold at which T is
calculated (mmHg).
B
Timescale Button
Press the button to cycle through the available
time scales (3 hr, 12 hr, 24 hr, 48 hr, 6 days).
Figure 8: Mean ICP Histogram Screen
C
T Thresholds
D
Histogram
12
Dashed Lines indicate T Thresholds.
Distribution of mean ICP within the selected time
window.