TROUBLESHOOTING
System Messages
System Message
Battery not detected.
Power down monitor and attach
battery.
Date and time changed!
Remove sensor to prevent data
deletion, or acknowledge to
continue.
Low Battery! 15 minutes left.
Connect monitor to power supply.
New Sensor Detected!
Previous sensor data will be deleted.
Disconnect new sensor now to stop
deletion.
No USB device recognized.
Patient Monitor disconnected.
Screen shot memory full.
The oldest image will be replaced.
Sensor disconnected from the cable.
Sensor or extension cable failure!
Disconnect and replace cable or
sensor.
Sensor implanted or not still.
Cause
Battery not connected.
System date and time
modified while the
sensor was temporarily
disconnected.
Approximately 15
minutes of battery charge
remaining.
Stored data will be deleted
when a new sensor is
detected.
Problem with the
connection or the
compatibility of the USB
memory device.
Connection with the
Patient Monitor lost during
the Patient Monitor Setup.
Maximum number of screen
shots has been reached.
ICP sensor disconnected
from the extension cable.
Extension remains
attached to the monitor.
System has detected an
issue with the sensor or
extension cable.
System has detected
pulsatility in the pressure
sensor reading. This
indicates the sensor is
implanted or moving prior
to the zeroing process.
25
Recommendation
• If battery is connected, replace the battery. Refer to
Replacing the Battery section.
• If problem persists, contact Technical Support.
• Only update date and time before connecting a new sensor.
• Download stored data to a USB device prior to
reconnecting the sensor.
• Connect monitor to Power Supply to recharge the battery.
• Download data prior to connecting a new sensor. Refer to
Data Download section.
• Reconnect the USB memory device and try the operation
again.
• Make sure the USB memory device is plugged in all the way.
• Make sure the USB memory device meets the minimum
specifications. Refer to Data Download section.
• Use a different USB device with no files stored in memory.
• Reboot monitor and try again.
• Reconnect Patient Monitor and perform the calibration
process.
• Make sure interface cable is connected properly.
• Replace the interface cable.
• Make sure Patient Monitor is turned on.
• Download all screenshot to a USB memory device. Refer to
Data Download section.
• Disconnecting the cable from the monitor will not trigger
this message.
• If message appears while sensor is still attached, sensor or
the cable may be damaged.
• Make sure the sensor is connected correctly into the
extension cable.
• Make sure the extension cable is connected correctly to the
monitor.
• Make sure to use a CereLink ICP sensor (refer to
Recommended Accessories).
• If problem persists, replace extension cable.
• If problem persists, replace ICP Sensor.
• If problem persists, contact Technical Support.
• An implanted sensor will not be zeroed.
• Place sensor in a shallow pool of sterile water and make
sure it is not moving.
• If problem persists, replace extension cable.
• If problem persists, replace ICP Sensor.
• If problem persists, contact Technical Support.
EN - ENGLISH