Section 11: Specifications
Instrument
Number of Channels:
Timing Range (Seconds):
Actuator Heat Block Temperature:
Range:
Resolution:
Accuracy:
Electrical
Note: Technical data, features, and options referenced in this manual are based on
the latest information available at the time of printing. Medtronic reserves the right
to change specifications without notice.
Classification (IEC 61010):
Power:
Voltage:
Frequency:
Maximum current:
Fuse:
Power cord:
Length:
Type:
Safety Standards:
EMC Standards:
Physical Dimensions:
Depth:
Width:
Height:
Weight:
Environmental:
Operating temperature:
Operating humidity:
Storage temperature:
Storage humidity:
Instrument use:
Altitude:
Mains supply voltage fluctuation:
Transient overvoltages:
Pollution degree:
®
ACT Plus
Operator's Manual
2
6 - 999
37.0°C ± 2°C
0.1°C
± 0.5°C
Class I, IPX0, Continuous operation
100 - 240 V~ Single phase
50 - 60 Hz
1.0 A (100 - 240)
T2A 250 V
2 wires plus a ground (earth) connector
3.1 m (10 ft.)
3-prong hospital grade (USA only)
IEC 61010-1: 2001
CSA C22.2 NO. 61010-1
IEC 60601-1-2: 2007
33 cm (13.0 in)
20 cm (8.0 in)
27 cm (11.0 in)
5.22 kg (11.5 lb)
14°C to 32°C (57°F to 90°F)
10% to 90%, noncondensing
0°C to 49°C (32°F to 120°F)
5% to 90%, noncondensing
Indoor
≤ 2000 meters
Not to exceed ± 10% of nominal voltage
Category II
2