Alerts

Bar code scan timeout.
> 10% spread error.
> 12% spread error.
Error - File not found.
52 English
When changing the password for the QC
Manager, the Confirm Password screen did
not match the Change Password screen.
Reenter the same password in the Confirm
Password screen that was entered in the
Change Password screen.
The QC for this cartridge has passed its
interval. Perform QC tests for this cartridge.
The test stopped before clots were detected
in both channels. Repeat test.
The QC Manager is trying to download User
IDs to the floppy disk or USB storage device,
but no User IDs are in the memory. Add User
IDs to the memory and download to the
floppy drive or USB storage device.
The QC Manager is trying to upload User
IDs from the floppy disk or USB storage
device, but there are no User IDs in the file.
Download User IDs from an ACT Plus
instrument or an ACT Plus
Manager (EDM) to the floppy disk or USB
storage device, and use that floppy disk or
USB storage device to upload them.
The QC Manager has set User Lockout to
ON so that a User ID is required to run a
patient test. Enter a valid User ID in the
Enter ID screen.
A bar code input sequence was detected,
but it did not complete correctly. Rescan the
bar code.
The channels of the controls tested are not
within 10%. Repeat the control test.
The channels of the controls tested are not
within 12%. Repeat the control test.
®
The ACT Plus
instrument is trying to read a
file on a floppy disk or USB storage device,
but cannot find it.
ACT Plus® Operator's Manual
®
®
External Data