Descargar Imprimir esta página

Sartorius Gold & Carat II Line GL822I-2 Serie Instrucciones De Manejo página 16

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 129
System Settings
7
System Settings
7.1
Performing System Settings
Default settings can be adjusted for the device and the
applications in order to align with the ambient condi-
tions and individual operating requirements.
The following settings are necessary to operate the
device together with connected components:
— Set up the communication of the connected devices
— Set up additional components
The following settings are recommended to set up the
device:
— Set the menu language
— Set the date and time
— Set the calibration and adjustment
Procedure
Press the [Menu] button.
t
To adjust settings: Open the desired menu.
t
Select and confirm the desired parameter (parame-
t
ters, see Chapter "7.3 Parameter List", page 17).
Exit the menu.
t
7.2
Setting the Calibration and
Adjustment
7.2.1
Setting Internal Calibration and
Adjustment (Only Model I-2x)
The following functions can be set for the internal
calibration and adjustment:
— Internal calibration with automatic start of the
adjustment.
— Internal calibration with manual start of the
adjustment.
16
Operating Instructions Gold&Carat II Line
Procedure
Open the "SETUP"/"BALANCE" menu. Call up the
t
"CAL.JUST." parameter and select the value "CAL.
INT.".
If the calibration function needs to be set with
t
subsequent automatic adjustment: In the "SETUP"/
"BALANCE" menu, for the "CAL.SEQ." parameter,
select the "ADJUST" setting value.
If the calibration function needs to be set without
t
subsequent automatic adjustment: In the "SETUP"/
"BALANCE" menu, for the "CAL.SEQ." parameter,
select the "CAL./ADJ." setting value.
7.2.2
Setting the External Calibration and
Adjustment
The following functions can be set for the external
calibration and adjustment:
— External calibration with automatic start of the
adjustment.
— External calibration with manual start of the
adjustment.
If this relates to a conformity-assessed de-
vice in legal metrology: External adjustment
is not possible for accuracy class II devices.
Procedure
Open the "SETUP"/"BALANCE" menu.
t
Call up the "CAL.JUST." parameter.
t
Select the value "CAL.EXT".
t
If the calibration function needs to be set with subse-
t
quent automatic adjustment: Select the "ADJUST"
calibration value for the "CAL.SEQ." parameter.
If the calibration function needs to be set without
t
subsequent automatic adjustment: Select the
"CAL-ADJUST" calibration value for the "CAL.SEQ."
parameter.

Publicidad

loading